This is to inform all concerned healthcare professionals and establishments that the Voluntary Recall of affected lots/batches of Hurricane RX Biliary Balloon Dilatation Catheter manufactured by Boston Scientific Limited and imported by Boston Scientific Philippines, Inc. as shown in the table below is completed and hereby closed by the Food and Drug Administration (FDA).

Product Description

Material / Reference No. Lot / Batch

Hurricane RX Biliary Balloon Dilatation Catheter 8MM 4CM

M00545940

26473775, 27110099, 26406534

Hurricane RX Biliary Balloon Dilatation Catheter 10MM 4CM

M00545960

27065900, 26677002, 26651709

Hurricane RX Biliary Balloon Dilatation Catheter 6MM 4CM

M00545920

27100240

Hurricane RX Biliary Balloon Dilatation Catheter 4MM 4CM M00545900

26087729, 26087720, 26240367, 26207977

 

As stated in the FDA Advisory No. 2022-1021 dated 04 April 2022, Boston Scientific Philippines, Inc. has conducted the voluntary recall of the aforementioned product in response to an increase of complaints reported for pinholes in the balloon. The user may notice that the balloon either rapidly loses pressure or fails to gain or maintain pressure. The most common potential consequence that is reasonably foreseeable to occur is a prolonged procedure while the device is exchanged. If a prolonged procedure were to occur, it is not anticipated to result in a clinically significant delay in the procedure. Furthermore, there have been no reports of injuries related to this recall to date.

After due and thorough evaluation of the submitted documents by Boston Scientific Philippines, Inc., FDA has determined that reasonable efforts have been made to recall and properly returned to the source country the affected product batches in accordance with FDA Circular No. 2016-012, known as the Guidelines on Product Recall.

The issuance of this advisory shall not in any manner preclude this Office from issuing subsequent orders it may deem necessary and appropriate, should there be subsequent findings of any violation of existing FDA laws, rules and regulations.

All FDA Regional Field Offices and Regulatory Enforcement Units in coordination with law enforcement agencies and Local Government Units are requested to monitor and seize the cited product lots if still found available in the market.

For more information and inquiries, kindly contact the FDA Center for Device Regulation, Radiation Health and Research through e-mail at [email protected], or call (02) 8857-1900 local 8301.

Dissemination of this advisory to all concerned is hereby requested.

Attachments